Description
Features
- CPAP ventilation mode keeps children spontaneous breath and reduces children breathing
work, continue to provide CPAP mode of constant flow and limited pressure for new born baby. - Delivery air/ O2 mixture with properly adjusted O2 concentration and continuous positive
pressure to premature baby, new born baby and child, which will highly improve the safety
and efficacy of oxygen therapy; - Equipped with built-in oxygen/air blender, ensuring O2 concentration adjustment to be stable.
- High quality oxygen sensor can monitor real-time O2 concentration, alarm over limitation,
ensure safety of breathing O2. - Reliable CPAP control valve and advanced pressure monitoring system improve CPAP adjustment precision.
- 3’’ display
Working condition :
- Air source O2, AIR
- Air source pressure 280kPa~600kPa
- Power supply~220V±22V,50HZ with air compressor
100-240V,50/60HZ (without air compressor) - Input power 900VA(with air compressor)
250VA (without air compressor)
Technical Parameters of CPAP Infant Ventilator
Gas supply
- Gas source O 2 (For medical use) Air(For medical use)
- Rated working pressure 350 kPa
- Output pressure of regulator 350 kPa±50 kPa (300 kPa ~400 kPa )
Power supply
- Voltage ~220 V±22 V
- Frequency 50 HZ±1 HZ
- Input power 900A
——Main unit 25
——Air compressor 675 W - Internal power DC12V 4Ah
- Fuse
——AC power for main unit 5T T1AL 250
——Battery 5T T2AL 250V
——AC power for air compressor 5T T6.3AL 250V
——AC power for machine 5T T10AL 250V
Parameter adjustment
- FiO2
——Adjusting range 21 %~100 %
——Accuracy ± 15 % - CPAP adjusting range 0 cmH 2 O ~15 cmH2 O
- Continuous flow adjusting range 2 L/min ~20 L/min
- Continuous flow adjusting range for manual 20 L /min±5L/min
Parameter monitoring
- FiO2
——Adjusting range 15 %~100 %
——Accuracy ± 5 % - CPAP/airway pressure
——Monitoring range 0 cmH2 O ~60 cmH2O
——Accuracy ±(2%fullscale +4 % actual value) - Continuous flow
——Monitoring range 2 L/min~20 L/min
——Accuracy ≥3 L/min:± 20 % - Waveform display:
——airway pressure—time waveform
Technical specification of compressor
- Rated output pressure: 0.35 MPa±0.05 MPa(350 kPa±50 kPa)
- Continuous output flow ≥25 L/min when output pressure is 0.35MPa
- Peak flow: ≥120 L/min
- Dew point temperature ≤-10°Cwhen Atmospheric pressure is 860 hPa ~1060hPa
- Alarm for high heat: There is sound alarm and light alarm when the compressed pump temperature is over 85 °C±10 °C
- Noise ≤50 dB(A)
Reviews